Open Care System 5100/5000 (OCS 5100-5000 Series)

/Open Care System 5100/5000 (OCS 5100-5000 Series)
Open Care System 5100/5000 (OCS 5100-5000 Series)2019-03-16T08:43:43+00:00

FEATURES

  • Micro Processor based servo-controlled electronic controller
  • Source box made with high quality F.R . grade fiber material for
    extra safety from electric shocks and corrosion
  • Newly designed oval shape baby tray with four sides collapsible
    clear acrylic panels,head up & down tilting facility for baby tray
  • Baby Matress density aprox. 21-25 Kg./m3 with removable
    washable cover.
  • LED Examination Lamp
  • German make ceramic heater with parabollc reflector
  • Maintains uniform temp across the mattress with swlvellng source box easy to take x-ray  procedures
  • Skin/Air/ Pre warm/ Manual control with LCD Screen
  • Thermistor Based Skin/Air Sensor
  • Visible LED Display for Skin/Air Temperature/Set Temperature & % of heater power output
  • LED Indicator for selected mode
  • Bright back-light LCD Screen for alarm messages,Selected mode,timer Dlsplay & SkinTemp. in  C° & F°
  • IntegratedTimer 1-59Min With count-up & count -Down Feature & APGAR
  • Temp.range -Skin-34 to 38°C (User Pre-settable),Air -Room Temp.to 39°C
  • Accuracy:+/- 0.1C°,Resolution: 0.1°C
  • If the Heater Power is 100% for more than 15 mins. heating comes down to 25%( Check Sensor Position)
  • Heater is unplug or faulty-Heater open.
  • Heater Power Drive fall -Check Power Drive
  • Manual Mode-User is allowed to set the heater power from
    25%-100% with Increment of 5%. Tlme Limit 15 min.
  • Pre-warm Mode• User Is not allowed to set heater power more
    than 25%. Adjustable Heater Power 0-25% with increment of 5% Time limit 30 minutes
  • Heater Out put -0-100% inincrements of 5%
  • Control Unit- Auto Over- temp.Cutoff  With audio-visual alarms
    according to Preset Temp to avoid Over Heating
  • L.C.D. Display  Reports: systems errors,sensor failure alarm and visual
    message on LCD For High/Low,Over /Under Temp.for skin & air,Heater Failure,
  • Power Control device Failure. Vlsual /Audio Alarm for Power failure
  • Latest high bright super flux LED Top surface phototherapy with
    digital timer, integrated with heater source( Only In OCS 5100)
  • Facility to Incorporate Under Surface Phototherapy.
  • Keypad lock Facility
  • OperatfngVolatage:220V-2.40V 50HzAC
  • 2/3 Drawer  (Optional)
  • Peripheral temperature Display (Optional)
    Resuscitation Attachment for OCS 5100/5000( Open Care system)
  • Infant Resuscitator will safely inflate the Baby’s lungs and provide optimum oxygenation by delivering consistent PlP & PEEP/CPAP with each breathe avoiding the risks associated with under or over inflation of uncontrolled pressures.
  • The patient T-Piece can connect to Neonatal masks or Endotraceal tubes directly.
  • Different sizes of face mask which provides flexible in usage for different babies
  • lnspiratory high pressure safety cut-off relief setting
  • Peak inspiratory Pressure setting(PIP)
  • Positive and expiratory pressure setting(PEEP)
  • Works at flow range from 5L/min. to 15L/min.
  • Medical grade monometer to monitor PIP & PEEP delivered.
  • Inlets for oxygen and air with suitable dehumidifier at air inlet.
  • Integrated slow suction unit with vacuum regulator,vacuum gauge and auto
    cleavable suction jar and tubings.
  • Power requirement :220V-2.40V 50HzAC

Enquiry